ICON-X

Ex-certified configuration kit for INOR transmitters

ICON-X is a complete kit for configuration of all INOR PC programmable transmitters. Communication with the connected transmitters is automatically established and configuration is done with the ConSoft software. ICON-X has built-in protection for Ex-certified transmitters. Configuration software ConSoft and installation guide can be downloaded from the download center.

Cables for connecting all transmitters

ICON-X includes all the cables you need to connect to all INOR PC programmable transmitters*.

Protection against supply of energy in an explosive atmosphere

The ICON-X kit has built-in protection for the transmitter to ensure that no energy is supplied that could lead to an increased risk of explosion. ICON-X is Ex-certified, which allows the transmitter to be configured in a safe area with the temperature sensor still installed in an explosive atmosphere.

Optimised for work on the field

The configuration kit is portable and easy to use in the field. The new case made of hard plastic is designed to cope with tough handling.

The configuration kit consists of:

  • USB interface
  • Cables (Between transmitter, USB interface and PC)
  • Hard plastic case

 

* To configure OEM202 variants or Compact sensors there is a need for a special cable and adapters, not included in the standard kit, and needs to be ordered separately.

Category: .

Technical Specification

PC to USB InterfaceUSB cable type A to type B
USB Interface to transmitterChangeable adapter cables
Power SupplyPC´s USB port, 5 VDC, 74 mA max
Operation temperature0 to 50 °C / 32 to 122 °F
Storage temperature-20 to 70 °C / -4 to 140 °F
Humidity (non-condensing)0 to 90 %RH
Galvanic isolation1500 VAC
System requirements
Windows32- or 64-bit edition of Windows
XP (SP2+) / Vista / 7 / 8 / 8.1 / 10
Free hard disk space185 MB
PC Port1x USB 1.1 or higher port type A
EX data
0539 II (1)G [Ex ia Ga] IICUo : 9.4 V
KIWA 16ATEX0011XIo : 96 mA
IECEx KIWA 16.0005XPo : 0.68 W
USA FM18US0117XUm : 250 V AC/DC
Canada FM18CA0056X
AIS CL I, Div 1 and 2, GP A-D,
CL I, Zn 0,1 and 2 [AEx/Ex ia Ga] IIC

Documents

Request a quote

  • For example approvals and configuration
  • This field is for validation purposes and should be left unchanged.